Home » Shop » hot sale » Bard Silastic Two-Way Standard Specialty Foley Catheter With 30cc Balloon Capacity

Bard Silastic Two-Way Standard Specialty Foley Catheter With 30cc Balloon Capacity

Original price was: $14.76.Current price is: $8.89.

Sale!
SKU: 347 Category:

Bard Silastic Two-Way Foley Catheter Features

  • This indwelling catheter comes with 30cc Balloon Capacity
  • features silicone coating for less friction
  • Firm and flexible
  • Lubrication: Non-stick silicone coating

What to buy with Bard Silastic 2-Way Specialty Foley Catheter

  • Bard Bedside Urine Drainage Bag With Anti-Reflux Device
  • Bard Catheter Plug And Cap
  • Deluxe Catheter Bag

Inflation and Deflation Instructions for Two-Way Foley Catheter

  • Catheter Inflation:
    • Ensure that the Bard Foley catheter balloon is positioned well within the patients bladder. Slowly, with a gentle, constant force, inflate the Bard Foley catheter balloon with the volume prescribed on the package
    • Note that the 5cc balloon must be inflated with between 9cc to 10cc of sterile water
    • Improperly inflated Bard Foley catheter balloons may cause drainage and deflation difficulties
    • Using fluid other than sterile water may cause the balloon not to empty properly, especially after long dwell times
  • Catheter Deflation:
    • Select a luer slip syringe
    • Slide the plunger of the syringe up and down the barrel of the syringe several times to loosen it up
    • Compress the plunger all the way and then pull back the plunger slightly so that it does not adhere to the front of the syringe barrel
    • Gently insert the syringe in the catheter valve
    • Do not use more force than is required to make the syringe stick in the valve
    • Allow the pressure within the balloon to force the plunger back and fill the syringe with water
    • If you notice slow or no deflation, re-seat the syringe gently. Once again, allow the balloon to deflate slowly on its own
    • If the balloon does not deflate, reposition the patient.
    • Ensure that the catheter is not in traction, the proximal end of the catheter is not compressed within the bladder neck
    • Ensure that urine flows freely
    • Attempt to deflate the balloon by using the pressure in the balloon to force water into the syringe as described above
    • If the balloon stills fails to deflate, apply very gentle slow aspiration. Aspiration that is too rapid, or too forceful, may cause the inflation lumen within the Foley catheter to collapse
    • If permitted by hospital protocol, the valve arm may be severed
    • If this fails, contact an adequately trained professional for assistance, as directed by hospital protocol

Warnings:

  • On catheter, do not use ointments or lubricants having a petrolatum base. They will damage latex and may cause balloon to burst.
  • After use, this product may be a potential biohazard. Handle and dispose of in accordance with accepted medical practice and applicable local, state and federal laws and regulations.
  • Visually inspect the product for any imperfections or surface deterioration prior to use. Use Luer tip syringe to inflate with stated mL of sterile water OR for pre-filled products, remove clip and squeeze reservoir to inflate with stated mL of sterile water.
  • Store catheters at room temperature away from direct exposure to light, preferably in the original box